Multiparameter meters measure pH mV Temperature, Conductivity, Salinity, TDS, Dissolved Oxygen

Two channels - one can be used for pH mV and the other channel can be used at the same time for either Dissolved Oxygen OR Conductivity

Features of the C5010 Series
Multi probe electrochemistry meters

  • pH: 0...14 pH
  • mV: ±1000 mV
  • Conductivity: 0...100 mS/cm
  • Dissolved oxygen: 0...20 mg/l /0...200%
  • Temperature: 0...100°C
  • One pH/mV channel
  • One conductivity/oxygen channel
  • One temperature channel

Features of the C5020:

  • Same as C5010 above +
  • Conductivity: 0...1000 mS/cm
  • Salinity: 0...70 ppt
  • TDS: 0...100 g/l
  • Memory 300 readings

pH - Automatic calibration with one up to three buffers out of eleven pre-program med pH buffers. Reads pH with 0.01 pH resolution.
mV- Features mV calibration for accurate ORP mea su rements. Reads potentials with 1 mV resolution.
Conductivity- Use a 1 cm-1 electrode (standard) to measure from 0.1 µS/cm to 100 mS/cm.
Use a 0.1 cm-1 electrode to measure from 0.01 µS/cm to 10 mS/cm (C5020 only).
Use a 10 cm-1 electrode to measure from 1 µS/cm to 1000 mS/cm (C5020 only).
Automatically selects cor rect range and frequency.
Selectable reference temperature: 20° or 25°C.
Automatic calibration with any of three pre pro grammed standards (0.01, 0.1 or 1 M KCl)
Dissolved oxygen- Operates with a galvanic oxygen electrode requiring no polarisation time and no zero calibration.
Reads dissolved oxygen with 0.01 mg/l or 0.1% resolution.
Rapid air calibration.
Temperature- Reads temperatures with 0.1°C resolution. Manual or automatic temperature compensation. Calibrates temperature probe for quality measurements.
Inputs - One common input for pH and mV.
One common input for conductivity and dissolved oxygen.
One input for a Pt1000 automatic temperature probe.
Low voltage DC input for a mains adaptor.
Display - Bright LCD screen for better readability. A white backlight automatically illuminates when operated on the mains. Stability indicator prompts the user when readings should be taken.
The interactive LCD screen provides step by step instructions in the language of your choice (English, Dutch, French, German). Shows a GLP report on the LCD screen.
Data-logging- Storage memory for 300 values including temperature. Cabinet- Robust dust and splash-proof cabinet.
Special features Three year warranty.
Mains and rechargeable battery

 

Specifications C5010C5020
pH Range 0...14 pH 0...14 pH
Resolution 0.01 pH 0.01 pH
Accuracy 0.2% ± 1 digit 0.2% ± 1 digit
Calibration 1...3 points 1...3 points
Buffers 11 pre-programmed 11 pre-programmed
Temperature compensation 0...100°C 0...100°C
ISO-pH 6...8 pH 6...8 pH
Slope 80...120% 80...120%
mV Range ±1000 mV ±1000 mV
Resolution 1 mV 1 mV
Accuracy 0.2% ± 1 digit 0.2% ± 1 digit
Calibration 1 point 1 point
Conductivity Range (cc dependent) 0...100 mS/cm 0...100 mS/cm
Resolution (cc dependent) 0.1 µS/cm 0.1 µS/cm
Accuracy 1% f.s. of range 1% f.s. of range
Calibration 1 point 1 point
Standards 0.01/0.1M KCl 0.01/0.1M KCl
Cell constant (cc) 1 cm-1 ± 30% 0.1/1/10 cm-1 ± 30%
Temperature compensation 0...100°C 0...100°C
Reference Temperature 20° or 25°C 20° or 25°C
Temperature Coefficient natural waters (EN27888) natural waters (EN27888)
Salinity Range - 0 ... 70 ppt
Resolution - 0.1 ppt
TDS Range - 0 ... 100 g/l
Resolution - 0.1 mg/l
Dissolved Oxygen Range (cc dependent) 0...20 mg/l (0...200%) 0...20 mg/l (0...200%)
Resolution (cc dependent) 0.01 mg/l (0.1%) 0.01 mg/l (0.1%)
Accuracy 1% ± 1 digit 1% ± 1 digit
Calibration 1 point 1 point
Temp. Compensation 0...50°C 0...50°C
Salinity Compensation 0...40 ppt 0...40 ppt
Air Pressure Compensation 800...1200 hPa 800...1200 hPa
Temperature Range 0 ... 100 ºC 0 ... 100 ºC
Resolution 0.1ºC 0.1ºC
Accuracy 0.5°C 0.5°C
Calibration 1 point 1 point
Inputs pH/mV BNC, 1012 O BNC, 1012 O
Conductivity/ DO BNC BNC
Temp. 2 banana, for Pt1000 2 banana, for Pt1000
Storage Memory Data Sets - 300
Display LCD with white backlight 122x32 pixels 122x32 pixels
Ambient Conditions Temperature 0...40ºC 0...40ºC
Humidity 0...95%, non condensing 0...95%, non condensing
Power Supply Mains 210...250 VAC, 50/60 Hz 210...250 VAC, 50/60 Hz
Low Voltage 9...15 VDC 9...15 VDC
Batteries 4x1.2 V, NiMH 4x1.2 V, NiMH
Dimensions WxDxH 10x24x4 cm 10x20x4 cm
Weight meter 350g 350g